NON-CONTACT ELECTRICAL SUPPLY OF IMPLANTS

Володимир Петрович Олійник, Дар’я Вікторівна Теличко

Abstract


The issues of long-term implants using in the human body are considered. Particular attention is paid to the problem of contactless supply of power to artificial organs with significant energy consumption. Using non-contact energy supply to a fully implanted technical device reduces the risk of infections getting into the patient's body as compared to the use of a power line that is output through internal tissues to the skin surface. The well-known developments of artificial heart apparatus were used as the object of analysis. Their main technical characteristics are considered. The maximum power consumption of these devices is about 20 watts, including models AbioCor. The design of the AbioCor was produced as a fully implanted contactless power unit, but the project was not completed. Although the principle of contactless power supply is undoubtedly an innovative step in the development of implant designs. As a result of the technical implementation of this principle, the following was obtained: it is advisable to use an alternating magnetic field with an inductive coupling between an external and an internal inductor for non-contact energy supply of implants; the use of frequencies of an alternating magnetic field of ~ 100 kHz makes it possible to obtain a compromise solution between the required depth of penetration into biotissues, insignificant heat formation in them and the size and mass of inductors. Under such conditions, the location in the body of the receiving inductor can be determined by medical recommendations. Evaluation experiments were carried out to determine the effective scheme of inclusion of the transmitting and receiving circuits, and the effect on the transmission coefficient of the electrophysical properties of the environment between them. It is shown that for working out of technical solutions of hardware realization of contactless power supply, implantation elements can be placed in saline solution as a model of bioenvironment. It is recommended to use serial element base and circuit solutions used for contactless power supply of radio electronic devices. Also, a quantitative assessment of the additional heat load of the implant on the human body was carried out, since part of the energy of the apparatus and its work, in accordance with the efficiency, inevitably goes into the heat eventually. A simplified calculation of the impact of implant energy losses on the thermal state of the body showed that overheating by 1 °C would be achieved provided the complete insulation of the area, on average, 8.6 hours. This proves the non-critical overheating of the body with implants, and allows the maintenance of a stable body temperature by physiological mechanisms.

Keywords


artificial organs; implants; non-contact power supply; electromagnetic induction; power; thermal regime

References


Petuhov, D. S., Selishev, S. V., Telyshev, D. V. Perspektivy razvitiya tehnologij polnoj zameny funkcii serdca s pomoshyu mehanicheskih sistem podderzhki krovoobrasheniya [Prospects for the development of technologies for the complete replacement of cardiac function using mechanical circulatory support systems]. Medicinskaya tehnika – Medical equipment, 2015, no. 293, pp. 5-8.

Klark, Dzhon V. ml., Nyumen, Majkl R., Olson, Valter H. i dr. Medicinskie pribory: Razrabotka i primenenie [Medical Devices: Development and Application]. Kiev, Medtorg Publ., 2004. 620 p.

Kiourti, A. Konstantina, S. N. A review of In-body Biotelemetry Devices: Implantable, Ingestible and Injectable. IEEE Trans. Biomed. Eng., 2017, no. 464, pp. 1422–1430.

Belyaev, L. V. Laboratorno-stendovye us-tanovki dlya kompleksnoj ocenki harakteristik apparatov vspomogatelnogo krovoobrasheniya i iskusstvennogo serdca [Laboratory and bench installations for a comprehensive assessment of the characteristics of assisted circulatory apparatus and artificial heart]. Biotekhnosfera – Biotechnosphere, 2011, no. 4 (16), pp. 51-56.

Davydov, B. I., Tihonchuk, V. S., Antipov, V. V. Biologicheskoe deystvie, normirovanie i zashchita ot elektromagnitnykh izlucheniy [Biological action, regulation and protection against electromagnetic radiation]. Moscow, Energoatomizdat Publ., 1984. 177.

Timanyuk, V. A., Zhivotova, E. N. Biofizika [Biophysics]. Kiev, ID «Professional» Publ., 2004. 704 p.

Khokle, R., Essello, K., Bokor, D. Design, Modeling, and Evaluation of the Eddy Current Sensor Deeply Implanted in the Human Body. Sensors, 2018, vol. 18, iss. 11, article 38888. DOI: 10.3390/s18113888.

Zhdanov, A. V., Belyaev, L. V., Kulikov, S. V., Kilasev, N. B., Drobyshev, A. A. Sovremennyj podhod k proektirovaniyu iskusstvennyh zheludochkov serdca na osnove CALS-tehnologij [The modern approach to designing of the artificial ventricles of heart based on CALS-technologies]. Sovremennye problemy nauki i obrazovaniya – Modern problems of science and education, 2012, no. 5. Available at: http://www.science-education.ru/105-7077 (аccessed 12.05.2019).

Narayanan, R. Katushki induktivnosti dlya sistem besprovodnoj zaryadki [Inductors for wireless charging systems] Sbornik tekhnicheskikh statey Würth Elektronik 2016–2019 gg. – Würth Elektronik 2016–2019 Technical Articles Collection, pp. 181-186. Available at: https://www.symmetron.ru/suppliers/wurth/files/pdf/wurth/articles.pdf (аccessed 12.05.2019).

Oliinyk, V. P., Kulish, S. M., Telichko, D. V., Voloshin, Yu. A. Vibir robochoyi chastoty pristroyiv dlya bezkontaktnogo elektrozhivlennya implantativ [Choosing the operating frequency of implants for contactless power supply]. Zbirnyk materialiv naukovo–praktichnoyi konferenciyi «Modern methods, innovation, and experience of practical application in the field of technical science» – Collection of materials of scientific-practical conference "Modern methods, innovation, and experience of practical application in the field of technical science", Poland, Radom 24-25 deсember 2017, pp. 40-44.




DOI: https://doi.org/10.32620/reks.2019.3.03

Refbacks

  • There are currently no refbacks.